Oscilloscope v2
Build, use, and understand your own oscilloscope
Since completing the OSCAR and OSCARET project in September 2026 and publishing the book Arduino, Tcl/Tk osciloscopio, the project has taken several important steps forward:
- addition of a spectrogram
- integration with Sigrok
Several visual themes have also been added, and the interface has been internationalized.
Arduino UNO Tcl/Tk Signal acquisition DHT CA3306 DSO183
I had originally planned to wait until the second book was published before releasing this new version. In that book, we will build more powerful versions and hack several commercial oscilloscopes 1. However, I believe these new features already make OSCAR and OSCARET worth releasing now.
Warning
Here you will find information about TDSO, the software I have developed, whose accompanying book has not yet been published. To learn the fundamentals (excluding the spectrogram and Sigrok integration), see OSCAR y OSCARET
Visual themes
OSCAR: an Arduino UNO and two wires
The simplest way to get started is probably also the most surprising.
All you need is:
- an Arduino UNO
- a USB cable to connect it to your computer
- two wires to connect the signal you want to measure
That is all.
The ATmega328P’s built-in ADC digitizes the signal, while the computer displays and analyzes it.
In this way, an Arduino UNO becomes a small digital oscilloscope controlled from your PC.
The basic idea
Arduino acquires the samples, while the computer provides the display, graphical interface, and much of the processing.
This keeps the hardware extremely simple by moving many functions into software.
What can you do?
Among other things, you can:
- display signals in real time
- export captures for later analysis
- measure voltages
- measure frequency and period
- use different trigger modes
- analyze acquired samples
- examine the signal spectrum
- examine a signal spectrogram
- decode various protocols (UART, DHT11, etc.) using Sigrok
Above all, you can experiment.
OSCAR is not intended to replace a professional oscilloscope. Its purpose is to show just how much can be achieved with very simple hardware and help you understand what really happens inside a digital measuring instrument.
An important limitation
If you connect signals directly to an Arduino UNO, their voltage must remain approximately within this range:
$$ 0\text{ V} \leq V_{in} \leq 5\text{ V} $$Arduino cannot directly accept negative voltages or voltages above its supply voltage.
You must therefore be careful about which signals you connect.
Nevertheless, it can be a surprisingly useful tool for digital signals, sensors, small electronic circuits, PWM, suitably conditioned low-level audio, and many laboratory experiments.
If you do not have a development environment, I have provided a ready-to-run executable: TDSO.
Going a little further
If you want to go a step further and measure AC/DC signals at different amplitudes, see OSCAR y OSCARET.
TDSO user guide with OSCAR
1. Getting started
1.1. Connect the Arduino UNO R3 to your computer
Connect the Arduino UNO via USB and power up the oscilloscope. Upload the firmware using the Arduino IDE.
The original firmware uses a serial connection at 115200 bit/s, 8-N-1.
1.2. Run TDSO
Launch the TDSO executable.
The program automatically detects the serial port of genuine Arduino boards and starts communicating without any configuration. If you are using a clone, you will need to specify the serial port manually by editing session.ini.
Tip
The program is designed for use with a genuine Arduino UNO.
You can also use a clone, but you must specify its serial port in session.ini, for example COM8.
I recommend using a genuine Arduino UNO because, in my experience, clones can have serial communication issues and may stop sending data to the program. I have never encountered that problem with a genuine board.
Warning
The screenshots show a prototype with more capabilities than OSCAR. The functionality demonstrated is exactly the same.
2. Adjusting the signal display
Before taking measurements, adjust the vertical scale and timebase so that the signal occupies a substantial portion of the display.
As a rule of thumb:
- adjust V/div to make good use of the available height
- adjust the timebase to display several periods of the signal
- select AC or DC depending on the signal you want to observe
- configure the trigger to obtain a stable waveform
The demonstration shows a signal of approximately 10 kHz and 2 V.
3. Interface language and appearance
TDSO lets you change the language and choose from different visual themes.
These settings do not affect data acquisition, but they let you tailor the interface to your preferences and make it easier to use in educational or laboratory settings.
4. Automatic measurements
TDSO can automatically calculate several properties of the acquired signal, including its frequency.
For reliable automatic measurements, the acquisition must contain enough information about the signal.
Important
An unsuitable timebase or sampling rate can cause measurement errors. If the signal appears overly compressed or too few samples are captured per period, review the acquisition settings.
In the demonstration, TDSO correctly measures a frequency close to 10 kHz.
5. Triggering: hardware and software
For everyday use of OSCAR, the Auto trigger mode (software triggering) is generally the best choice. However, decoding protocols with Sigrok requires the Normal trigger mode (firmware-based triggering).
5.1. Auto, Normal, and Single modes
You can choose different trigger modes depending on the type of capture.
| Mode | Recommended use |
|---|---|
| Auto | General observation and continuous operation. |
| Normal | Capture only when the trigger condition is met. |
| Single | Capture a single event and stop acquisition. |
Single mode is especially useful for precise manual measurements or examining a particular event.
6. Cursors and manual measurements
When an automatic measurement is not enough, or you want to verify it manually, TDSO provides measurement cursors.
You can move the cursors with the mouse and fine-tune their positions using the arrow keys.
This lets you measure time intervals and calculate a signal’s frequency manually.
For example, if the measured period is approximately:
$$ T = 100\ \mu\text{s} $$then the frequency is:
$$ f = \frac{1}{T} = 10\ \text{kHz} $$7. Spectrum analyzer
TDSO includes a spectrum view that shows the frequency content of a signal.
This tool is particularly useful for:
- identifying the fundamental frequency
- detecting harmonics
- comparing the spectral purity of different signals
- studying noise and distortion
- investigating sampling-related effects
7.1. Choosing the right acquisition settings
When viewing the spectrum, avoid placing the fundamental frequency too close to the left edge of the graph.
The acquisition settings should provide sufficient resolution while keeping the frequency of interest within a useful part of the spectrum.
Caution
Remember that the spectrum is limited by the Nyquist frequency. Components above half the sampling rate can cause aliasing.
8. Spectrum cursors
The spectrum view has its own cursors for precise analysis of frequency components.
In the demonstration:
- the left mouse button positions the first cursor
- the right mouse button positions the second cursor
- each cursor displays the frequency and magnitude of the selected component
Magnitude can be displayed in dBFS.
This makes it easy to identify the fundamental and its harmonics. For example, a square wave shows odd harmonics: the third, fifth, seventh, and so on.
9. Spectrogram
In addition to the conventional spectrum view, TDSO includes a spectrogram.
While a spectrum shows the frequency content of a single capture, a spectrogram shows how that content changes over time.
It is especially useful for signals whose spectral content changes over time.
10. Protocol decoding with Sigrok
One of TDSO’s most powerful features is its integration with Sigrok for decoding digital signals.
Important
To decode protocols with TDSO, you need to install sigrok-cli. Decoding DHT11 signals also requires a small modification to the AM230x decoder, explained below.
10.1. Installing sigrok-cli
Download and install sigrok-cli from the official Sigrok downloads page. On Windows, choose the sigrok-cli installer for your system. You do not need to install PulseView to follow this guide.
Once installed, locate sigrok-cli and make sure TDSO is using that installation.
10.2. Adapting the DHT11/AM230x decoder
The original Sigrok AM230x decoder expects to see the entire beginning of the sensor’s start pulse. With OSCAR, however, acquisition may begin after that pulse has already started because of the limited buffer length and trigger position. As a result, the decoder may reject an otherwise valid frame.
To address this, the decoder has been modified to accept an initial pulse that is already low when sampling begins, but only at the very start of the capture. All other timing checks remain unchanged.
Recommended installation procedure (Windows):
- Download the modified AM230x decoder for TDSO.
- Locate the
sigrok-cliinstallation directory and openshare\libsigrokdecode\decoders\am230x. - Back up the original
pd.pyfile, for example aspd.py.bak. - Copy the downloaded
pd.pyinto that folder, replacing the original. - Restart TDSO if it was running, then test DHT11 decoding.
The relative path within the sigrok-cli installation is:
Warning
Replace only the AM230x decoder’s pd.py file, not any other decoders. Keep the backup so you can restore the original. Reinstalling or updating sigrok-cli may overwrite the modification. This adaptation is provided for the version of sigrok-cli used with TDSO; compatibility with all future versions is not guaranteed.
If you prefer to review or apply the changes manually, the am230x.diff diff file is also available. Note: this diff compares the modified file against the original (the reverse of a typical update patch), so installing the supplied pd.py directly is recommended.
10.3. When to use hardware triggering
For protocol decoding, it is important to capture the entire event at a predictable position in the buffer.
In this case, hardware triggering may be necessary because you need to capture the actual transmitted sequence, not merely display a stable waveform.
11. Example: decoding a DHT11
The DHT11 uses a simple digital protocol to transmit humidity and temperature readings.
In the demonstration, the sensor’s digital output is connected directly to the oscilloscope, and TDSO is configured to capture the frame.
11.1. Configure the capture
Set the following:
- DC coupling
- a suitable vertical scale
- a timebase long enough to capture the entire frame
- hardware triggering on the appropriate edge
DHT11 communication begins with a relatively long start pulse generated by the MCU, followed by the sensor’s response and the data bits.
11.2. Select the decoder
In the Sigrok section:
- select the DHT11 protocol
- enable real-time decoding
- inspect the annotations displayed over the waveform
TDSO can display numerous internal protocol annotations. If you only want the final results, select the results annotation: am230x=results. See the Sigrok documentation for more information.
This allows the interface to display only values such as:
12. Example: decoding serial communication
TDSO can also use Sigrok decoders to interpret serial communication.
Before decoding, make sure the entire frame fits in the acquisition buffer.
If it does not fit:
- increase the timebase
- change the trigger position within the buffer
Once the complete frame has been captured:
- select the appropriate serial communication decoder
- configure its parameters
- enable real-time decoding
- check the decoded values
In the demonstration, humidity and temperature readings are transmitted periodically.
13. Recommended workflow
For general use of an Arduino UNO R3 with TDSO, follow this workflow:
flowchart TD
A[Connect and power on Arduino] --> B[Run TDSO]
B --> C{Port detected?}
C -->|Yes| D[Adjust V/div and timebase]
C -->|No| E[Select port manually]
E --> D
D --> F[Adjust trigger]
F --> G{Goal}
G -->|View signal| H[Software trigger]
G -->|Single event| I[Single]
G -->|Decode protocol| J[Hardware trigger]
H --> K[Measurements / cursors]
I --> K
J --> L[Sigrok]
K --> M[Spectrum / spectrogram if needed]14. Troubleshooting
15. Summary
TDSO turns an Arduino UNO R3 into a far more versatile instrument than its original interface alone would allow. Its most useful features include:
- oscilloscope control from a PC
- automatic serial port detection
- software triggering
- Auto, Normal, and Single modes
- cursors and manual measurements
- automatic measurements
- spectrum analyzer
- spectrum cursors with frequency and dBFS readouts
- spectrogram
- integration with Sigrok
- protocol decoding
- control over the trigger position within the buffer
Once you have mastered the basic controls, the spectrum, spectrogram, and decoding features make TDSO + OSCAR a particularly useful platform for learning, laboratory work, and analyzing small electronic systems.